Measurement-based deep venous thrombosis screening system

TitleMeasurement-based deep venous thrombosis screening system
Publication TypeJournal Article
Year of Publication2003
AuthorsGuerrero, J., S. E. Salcudean, J. A. McEwen, B. A. Masri, and S. Nicolaou
Secondary AuthorsEllis, R. E., and T. M. Peters
JournalMedical Image Computing and Computer-Assisted Intervention - MICCAI 2003, PT 1
Volume2878
Pagination214–221
ISSN0302-9743
Abstract

An experimental system and interface that indicate the likelihood of deep venous thrombosis using objective measures was developed, based on ultrasound image processing using a modified Star-Kalman algorithm and a sensorized ultrasound probe. Force, location and image data is used to assess a vessel segment for compression. A user interface displays the results using a 3-D representation. A tissue phantom was developed for testing and validation. Initial results with this phantom and healthy volunteers are presented.